ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
4-38 Section 4-4 - Image Testing
PRELIMINARY
Section 4-4
Image Testing
This section provides guidelines for the following image testing checks:
- 2D Mode (B Mode) Checks - see below
- ‘M Mode Checks” on page 4-39
- ‘Color Mode Checks” on page 4-40
- ‘PW/CW Doppler Mode Checks” on page 4-42
- ‘Tissue Velocity Imaging (TVI) Checks” on page 4-43
- ‘Color Mode Checks” on page 4-40
- ‘Probe/Connectors Check” on page 4-44
NOTE: For more details, refer to the information provided in the Vivid S5™ and Vivid S6™ User Manual.
4-4-1 2D Mode (B Mode) Checks
4-4-1-1 Introduction
The 2D Mode is the Vivid S5™/Vivid S6™ system’s default mode.
4-4-1-2 Preparations
1.) Connect one of the probes.
See ‘Connecting Probes” on page 3-71 for information about connecting the probes.
2.) Turn ON the scanner.
The 2D Mode window is displayed (default mode).
Figure 4-9 2D Mode Screen Example
Status Window
Control Menu