ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-86 Section 8-3 - Control Console Components Replacement Procedures
PRELIMINARY
8-4-10 LEDs and Soft Menu Board Replacement (New-type OP)
NOTE: The procedures described in this section are applicable to all Vivid S5 N / Vivid S6 N systems supplied
with the new-type Operator Panel (Figure 8-93).
NOTE: For systems with the older-type Operator Panel, refer to the instructions provided in “LEDs and Soft
Menu Board Replacement (Old-type OP)” on page 8-88.
8-4-10-1 Tools
Use the appropriate T8 Torx screwdriver, as indicated in the LEDs and Soft Menu Board replacement
procedure.
8-4-10-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-4-10-3 LEDs and Soft Menu Board Removal Procedure (New-type Operator Panel)
1.) Remove the Keyboard Assembly, as described in the “Keyboard Assembly Removal Procedure” on
page 8-39.
2.) Carefully place the Keyboard Assembly upside-down on a flat, clean, stable surface.
3.) Remove the Keyboard Bottom Cover, as described in the “Keyboard Bottom Cover Removal
Procedure” on page 8-59.
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to “Battery Removal Procedure” on page 8-116.