Chapter 8 - Replacement Procedures 8-193
PRELIMINARY
8-7-3 Lower Arm Replacement Procedure
8-7-3-1 Tools
Use the appropriate flat and Phillips-type screwdrivers as indicated in the Lower Arm replacement
procedure.
8-7-3-2 Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-7-3-3 Lower Arm Removal Procedure
1) Working from the front of the system, raise the console to the maximum up position.
2) Make sure that the console is aligned in the central position (not pulled to one side or the other).
3.) Loosen and remove the screw that is located at the base of the operator panel (below the
keyboard), as shown in Figure 8-242 below.
Figure 8-242 Removing Screw on Left Side of Lower Arm