10-6-2 Current Leakage Limits
The following limits are summarized for NFPA 99 (for USA) and IEC 60601-1 Medical Equipment Safety
Standards. These limits are GE standards and in some cases are lower than the above standards listed.
Table 10-13 Chassis Current Leakage Limits - Accessible Metal Surfaces
Country Normal Condition Open Ground Reverse Polarity Open Neutral
USA
N/A
0.3 mA 0.3 mA N/A
Other
0.1 mA
0.5 mA 0.5 mA 0.5 mA
Table 10-14 Type BF Applied Part Current Leakage Limits
- Non-Conductive (Floating) Surface and Cavity Probes
Country
Normal
Condition
Open
Ground
Reverse
Polarity
Open
Neutral
Mains
Applied
a
a. Single fault condition.
ALL
0.1 mA
0.5 mA 0.1 mA 0.5 mA 5.0 mA
Table 10-15 Type CF Applied Part Current Leakage Limits - Surgical Probes and ECG Connections
Country
Normal
Condition
Open
Ground
Reverse
Polarity
Open
Neutral
Mains
Applied
a
a. Single fault condition.
ALL
0.01 mA
0.05 mA 0.01 mA 0.05 mA 0.05 mA