GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-4 Section 8-2 - Cover Replacement Procedures
PRELIMINARY
8-2-2 Right Side Cover Replacement Procedure
NOTE: There are two types of Right Side Cover (supplied as part of the cover kit illustrated in Figure 9-1 on
page 9-3). The shorter cover is designed for use on Vivid™ S5 N and Vivid™ S6 N systems on which a
B & W printer (optional) is installed - to allow space for the printer front panel. The longer cover is for
use on systems without a B & W printer (designed to close the empty cavity) - see Figure 8-2.
Replacement of both types of cover are described below.
NOTE: The right side cover is fastened in position on the side of the system by ball and snap-lock securing
clips. The shorter cover has four; the longer cover has five (see the example in Figure 8-2).
8-2-2-1 Tools
None.
8-2-2-2 Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
Figure 8-2 Right Side Cover (Longer Type) - Front and Inside Views
Front View Inside View
(showing ball and snap-lock securing clips)