GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
4-24 Section 4-3 - Functional Checks
PRELIMINARY
6
From the Test ECG device window, set the resolution to
1 second and press the Open Device button.
The ECG signal is displayed in the Data area, as shown
in the example below:
7
Verify that a single clear QRS signal is displayed.
8
When done, trackball to the Exit button and press Set.
9
When prompted by the following message, select Yes or
No as applicable.
9
Press the <Esc> button on the keyboard to close the
Diagnostic Test window.
Table 4-11 ECG Test (cont’d)
Step Task Expected Result(s)