ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
7-60 Section 7-4 - Performing Front End (FE) Diagnostics
PRELIMINARY
7-4-4 TR 32 Diagnostic Tests (cont’d)
NOTE: The Rx test has two test reports (Test Signal Amplitude and Frequency) which are listed in the
Reports window on the right. By default, the Test Signal Amplitude report is displayed in the
Reports window below (Figure 7-34 "View Test Log - Rx Test Signal Amplitude" on page 7-59).
NOTE: Left and right arrows provided in the View Report window enable scrolling between the different
channels.
Left-clicking on a currently-selected channel marks a red square on the report graph, and also evokes
the display of the following information relative to that specific point: Average, Deviation (standard
deviation from the Average), Max, and Min.
The signal should be within the limits marked in the white area (the dotted line represents the Average),
as shown in Figure 7-34 "View Test Log - Rx Test Signal Amplitude" on page 7-59.
8) When done, trackball to the Close button and press Set to close the View Report window.
9) Select the second report (Frequency); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-30 "View Test
Log - Second Harmonic" on page 7-55.
10.)Use the tools provided to scroll through the channels and view the results, as previously described.
11) When done, trackball to the Close button and press Set to close the View Report window.
12) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
Figure 7-35 View Test Log - Rx Test Signal Frequency