Troubleshooting and Diagnostics 294
V4.2 LabChip GX User Manual PerkinElmer
RF Tag Test
Test Chip A required
Pressure Leak Test
Test Chip A required
Chip Temperature Test
Test Chip A required
Description Potential Cause of Failure
Read RF Tag Bad tag on chip. Try another chip with
known good RF Tag.
Problem with RF Tag reader. Contact
PerkinElmer Technical Support (see
page 3).
Description Potential Cause of Failure
Marker Pressure O-Ring
Leak
Pressure leak at O-rings. Clean or
replace the O-rings (see page 281)
and clean the chip cartridge.
Problem with pressure lines. Contact
PerkinElmer Technical Support (see
page 3).
Priming Pressure O-Ring
Leak
Sipping Pressure O-Ring
Leak
Description Potential Cause of Failure
Chip Heater Ramp Initial chip temperature not achieved in
time. Temperature ramp time not within
spec may indicate heater disconnected
or burned out. Contact PerkinElmer
Technical Support (see page 3).