Chapter 2 Before Use
2-24
To output end event register data, set the bit corresponding to the enable
register to 1.
The end event register can be set to 0 using the
*CLS
command.
The enable register of the end event register cannot be changed using
*CLS
.
Error Event Register
The meaning of each bit of the error event status register is listed in the
table below.
Table 2.6.3-4 Meaning of Error Event Status Register
Bit Explanation
7
Not used; always 0
6
Not used; always 0
5
Not used; always 0
4
Not used; always 0
3
Not used; always 0
2
Meas-Condition
Becomes 1 at mismatch between current measurement
condition parameters (Active trace measurement
conditions) and result measurement condition
parameters
1
Peak/Dip
Becomes 1 when level peak or dip not found when peak
or dip search executed.
0
RES-Uncal
Becomes 1 when resolution setting not appropriate for
sweep width and sample count.