CAPINTEC, INC. CAPTUS
®
4000e
10-12 LAB TESTS November 16
Figure 10-14
Figure 10-15
To begin counting, verify that there are no radioactive sources nearby and select the OK
button. To abort the measurement, select the Cancel button and the screen will return to
that shown Figure 10-12. To exit to Figure 10-2 Schilling Test Main Screen, select the
Exit button or press Alt+E. To begin the test again, select the Start Procedure button or
press Alt+S.
5. If the Pre-Dose Measurement option was set to YES in the setup, the screen prompts
you to measure the pre-dose urine sample (patient blank) as shown in Figure 10-16.
Figure 10-16
To measure patient blank, place the patient blank in the well and select the Yes button or
press the Y key. To skip this step, select the No button or press the N key.
• To skip the second repetition or Pre-Dose measurements, select the No button or
press the N key when prompted to do so.
6. The screen appears as in Figure 10-17, prompting you to measure the Co57 standard.