EasyManua.ls Logo

Capintec CAPTUS 4000e - Page 81

Capintec CAPTUS 4000e
370 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
4000e
August 22 QUALITY ASSURANCE 5-17
Figure 5-17 Quality Setup Screen
4. To change the default values in Chi-Square Test frame in Setup, click in the text boxes
and input the new values.
Note: The Counting Time can be set between 2 and 9,999 seconds. The Repetitions
can be set between 5 and 20.
5. Select the Done button or press Alt+D to save the edited information or select the
Cancel button or press Alt+C to cancel the changes and return to Figure 5-16 Chi-
Square Test Screen.
6. Select the Start Procedure button or press Alt+S to start the test. The screen appears as
shown in Figure 5-18.
Figure 5-18
Note: If calibration has not been performed for the current day on the selected detector,
you will be prompted with the screen as shown in Figure 5-19 If calibration is
performed now, the program will return to Figure 5-16 Chi-Square Test Screen
after calibration is complete.

Table of Contents