Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 20 Isolette® Infant Incubator (Model C2000) Service Manual
28.Perform the following to check the operation of Air Mode:
a. Close all access openings.
b. Allow the incubator to warm up to the air set temperature of
95 F (35 C).
The incubator takes less than one hour to warm up to the air set
temperature of 95 F (35 C).
Yes No
↓→Go to RAP 2.34 on page 2-99.
29.When the Air Temperature display stabilizes, the digital display
remains within 0.5 C of the set temperature for 15 minutes.
Yes No
↓→Go to RAP 2.35 on page 2-101.
30.Perform the following to check the Low Air Temperature alarm:
a. On a unit with a Series 00 hood assembly (D), slide the blue
slide access panel latches (J), and open the access panel (I)
(see figure 2-2 on page 2-11).
or
On a unit with a hood assembly (D) later than Series 00, rotate
the access panel pawl latch knobs (K), and open the access
panel (I).
b. Pivot the access panel (I) to the full-open position so it hangs
straight down.
c. Allow the unit to operate.
In less than 5 minutes, the Alarm indicators illuminate, the Low Air
Temperature message appears on the Trend display, and an alarm
sounds.
Yes No
↓→Go to RAP 2.36 on page 2-102.
31.Perform the following to check the Low Skin Temperature alarm:
a. Close the access panel (I), and secure it using its blue slide
access panel latches (J) or its access panel pawl latch knobs
(K).
b. Connect the skin probe to skin connector #1 on the sensor
module (F) (see figure 2-8 on page 2-19).
c. Place the skin probe 4" (10 cm) above the center of the mattress.
d. Select Skin Mode.