Function Checks
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 21
2
e. S et the skin set temperature to 95 F (35 C).
f. When the temperature stabilizes, position the skin probe outside
of the incubator.
g. Allow the unit to operate.
In less than 5 minutes, the Alarm indicators illuminate, the Low
Skin Temperature message appears in place of the Trend display,
and the alarm sounds.
Yes No
↓→Go to RAP 2.37 on page 2-103.
NOTE:
The sensor module is equipped to accept two skin probes. However,
when the second skin probe is connected to the sensor module in Skin
Mode, an alarm sounds, and the message Remove Skin 2 Probe
appears. To connect the second skin probe, first select Air Mode, the
controller then displays two temperatures.
32.Perform the following to check the Skin Probe Disconnect alarm:
a. Remove the skin probe from the 2 skin connector on the sensor
module (F) (see figure 2-8 on page 2-19).
b. Select Skin Mode.
c. Disconnect the skin probe from the 1 skin connector on the
sensor module (F).
The alarm sounds, the Skin Temperature display goes blank (--.-),
and the message Skin Probe Disconnect appears.
Yes No
↓→Go to RAP 2.38 on page 2-104.
33.Perform the following to check the maximum air temperature.
a. Press the Silence/Reset key to silence the Skin Probe
Disconnect alarm for 5 minutes.
b. Connect a skin probe to the 1 skin connector on the sensor
module (F) to return the unit to normal operation.
c. If necessary, select Skin Mode.
d. Select a skin temperature setpoint of greater than 99 F (37 C).
e. Place the skin probe outside of the incubator.
f. Allow the incubator to heat.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals