Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 22 Isolette® Infant Incubator (Model C2000) Service Manual
g. When the Skin Temperature alarm sounds, press the
Silence/Reset key.
The incubator’s temperature stays below 10 2.0 F (38. 9C), as
indicated by the Air Temperature display.
Yes No
↓→Go to RAP 2.39 on page 2-105.
34.The incubator is equipped with a scale assembly.
Yes No
↓→Go to step 38.
35.Perform the following to check the operation of the scale assembly:
a. Ensure that the mattress tray is level and not in Trendelenburg
or Reverse Trendelenburg.
b. If you desire a pounds/ounces display, refer to “Installation and
Set-up” on page 6-13.
c. Remove any objects from the mattress.
d. Select Display 2, and press the Weigh softkey.
e. Press the Zero softkey twice in succession.
The Weight display reads 0, and the Weight Sample bar begins to
fill.
Yes No
↓→Go to RAP 2.47 on page 2-116.
36.Place a weight of known value less than 15 lb (7 kg) on the
mattress. The correct weight appears, and when the Weight
Sample bar is filled, a beep sounds and the weight is locked into the
Weight display.
Yes No
↓→Go to RAP 2.47 on page 2-116.
37.Perform the following:
a. Press the Home key to return to Display 2.
b. Press the Weigh softkey.
The display again displays the value of the weight on the mattress.
Yes No
↓→Go to RAP 2.47 on page 2-116.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Isolette C2000 and is the answer not in the manual?

Dräger Isolette C2000 Specifications

General IconGeneral
BrandDräger
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals