Alarms
Chapter 2: Troubleshooting Procedures
Isolette® Infant Incubator (Model C2000) Service Manual Page 2 - 29
2
For an explanation of humidity alarms, refer to table 2-3 on page 2-29.
Table 2-3. Humidity Alarms
For an explanation of oxygen alarms, refer to table 2-4 on page 2-29.
Table 2-4. Oxygen Alarms
Connect Skin 1
Probe
The alarm message appears, an alarm
sounds, and the Alarm indicators flash to
indicate that the Skin 1 Probe was dis-
connected from the sensor module during
Air Mode and an attempt has been made
to enter Skin Mode.
Go to RAP 2.42
on page 2-109.
Alarm Description Action
Low Humidity The alarm message appears with an
audible alarm and flashing alarm indica-
tor to indicate that the humidity tray is
empty.
Go to RAP 2.40
on page 2-106.
Humidity Heater
Failure
The alarm message appears to inform
the user that the humidity system mal-
functioned.
Go to RAP 2.40
on page 2-106.
Alarm Description Action
Low Oxygen % The alarm message appears, an alarm
sounds, and the Alarm indicators flash to
indicate that the oxygen concentration
within the hood environment is below the
oxygen setpoint.
Go to RAP 2.4 on
page 2-56.
High Oxygen % The alarm message appears, an alarm
sounds, and the Alarm indicators flash to
indicate that the oxygen concentration
within the hood environment is above the
oxygen setpoint.
Go to RAP 2.5 on
page 2-58.
Oxygen Calibration
Required
The alarm message appears, an alarm
sounds, and the Alarm indicators flash to
indicate that the oxygen control system
requires calibration.
Go to RAP 2.43
on page 2-112.
Alarm Description Action