ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
3-180 Section 3-4 - Preparing for Installation
PRELIMINARY
5.) If required, access the System Configuration utility, as follows:
a.) Make sure the System Configuration tab is activated as shown in Figure 3-209.
b.) System Configuration read-only information is displayed (in table format) under the following
column headings: Item, Value, Minimum Requirements.
c.) Below this information, the eVue Operation Utility buttons enable the user to enable or
disable the eVue mode, as required. An additional button (Configure eVue) is used to perform
the eVue configuration procedure. THIS PROCEDURE FOR AUTHORIZED USERS ONLY?
Note: In order to enable eVue operation, it is necessary to first perform the eVue configuration
procedure as described in the following step.
d.) Configure eVue (if required), as follows:
e.) Click the Configure eVue button. The Configuration dialog box opens:
Figure 3-209 MPEGVue Patient Manager Window - System Configuration Tab