Chapter 7 - Diagnostics/Troubleshooting 7-93
PRELIMINARY
7-5-1 Accessing the Monitoring Diagnostic Test Options (cont’d)
11.)Check that the sub-test passed and that result values are within the pre-defined Max and Min limits,
as shown in Figure 7-61 "View Test Log - Full System Monitoring Test - Temperatures" on page 7-
93.
12.)Select the third report (Fans Speed); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-62 "View Test
Log - Full System Monitoring Test - Fans Speed" on page 7-94.
Figure 7-61 View Test Log - Full System Monitoring Test - Temperatures