GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-28 Section 8-3 - Control Console Components Replacement Procedures
PRELIMINARY
8-3-1 Monitor Replacement Procedure (cont’d)
7.) Using a screwdriver, release the DVI cable and the power cable connectors from the monitor
(loosening the two screws on each side of the cables to disconnect them) - see Figure 8-31 below.
8.) Loosen and remove the four Phillips screws that secure the monitor to the support column, as
indicated in Figure 8-31 below.
Do not let the screws fall while performing this step
.
9.) Using two hands, carefully lift the monitor away from the securing bracket on the support column to
remove it from the system.
10.)Carefully lie the monitor face-down on a clean, soft surface.
Figure 8-30 Disconnecting the Power and DVI Cables
Figure 8-31 Removing the Securing Screws