GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-31
PRELIMINARY
3.) Remove the roller-screw from the Alignment Block - Figure 8-34.
4.) To remove the alignment block, first loosen the Monitor Arm locking bolt a few turns to create some
freedom of movement for the Monitor Arm assembly so that it can be raised and lowered on the
upper arm securing pin - Figure 8-35.
The object is to create sufficient space to remove the alignment block.
5.) Using a flat screwdriver, lever the alignment block upwards so that it will slide away from the Monitor
Arm securing pin - Figure 8-35. Remove the alignment block.
6.) Remove the Monitor Arm locking bolt from the Monitor Arm securing pin.
7.) Remove the four washers (carefully note the order of removal for easy replacement when
assembling).
8.) Remove the lower Ground lead - Figure 8-35.
Figure 8-34 Roller-screw Located on the Alignment Block
Figure 8-35 Monitor Arm Assembly Locking Bolt
Remove Roller-Screw
Alignment Block
Loosen
Monitor Locking Bolt
Lever Alignment Block
(move upwards)