ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-43
PRELIMINARY
8-3-4 Keyboard Assembly Replacement Procedure (cont’d)
When the installation process is complete, the following screen is displayed:
14.)Press Set on the keyboard, eject the CD, and reboot the system.
15.)With the system up and running, press Alt + L.
The Front Panel Lights Setup dialog box opens - see Figure 8-53.
16.)Set the AlphaNUM Lights as follows:
a.) Select the sliding indicator and slide it to its extreme left position.
b.) With the slide indicator still selected, press the right arrow on the keyboard 3 times to get to
the position as shown in the above picture.
17.)For the remaining panel lights setup, set each of them about halfway on the scale to reach the
settings as shown in Figure 8-53.
18.)Press Save and Press exit to return to the default scanning screen.
19.)Proceed to perform the following functionality tests:
- “Alphanumeric Keyboard Test” on page 4-18
- “Extended Keyboard and Trackball Test” on page 4-18
- “Speakers Tests” on page 4-21
- “Light Detector Test for Vivid S6 N only” on page 4-22
- “USB Test” on page 4-37
- “Grounding Continuity” on page 10-23
- “Chassis Current Leakage Test” on page 10-24
Figure 8-53 Front Panel Lights Setup Dialog Box