GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-129
PRELIMINARY
13.)Plug the adaptor into the AC power outlet and verify that the power indicator light is ON.
14.)Plug the adaptor’s USB cable into the system and wait a few seconds.
15.)Open the Explorer window (eight partitions will now be shown, as shown in Figure 8-159 below).
NOTE: The partition picture shown in Figure 8-159, above is based on Application Software Ver. 10.1.0 Build 68.
This might not correspond exactly with the partition shown on your system - this depends on the
application software version currently installed on your system.
16.)Check the Total Size and Free Space shown for partition E - the third partition) and the Total Size
and Free Space of the second Archive partition in Figure 8-159, above.
17.)From the User partition (H: in Figure 8-159, above), copy the sub-directory Marlin into the User
partition D:.
When the message Overwrite? is displayed, click Yes.
Wait until the copying process is complete (time required will depend on the size of the database).
18.)Disable the SQL Service by selecting Start > Programs > Sybase SQL Anywhere 8 > Sybase
Central, as shown in Figure 8-160 below.
Figure 8-159 Explorer Window showing Eight Partitions