ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-144 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
8-6-6 ECG Module Assembly Replacement Procedure (cont’d)
11.)Proceed to perform the following functionality tests:
- 10-6-6 "Isolated Patient Lead (Source) Leakage – Lead-to-Ground" on page 10-26
- 10-6-7 "Isolated Patient Lead (Source) Leakage – Lead-to-Lead" on page 10-27
- 10-6-8 "Isolated Patient Lead (Sink) Leakage - Isolation Test" on page 10-27
- 10-6-4 "Grounding Continuity" on page 10-23
- 10-6-5 "Chassis Current Leakage Test" on page 10-24
- 7-2-4-2-1 "Calibration" on page 7-12
- 4-3-1-7 "ECG Test" on page 4-23