8-146 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
8-6-7 DC-to-DC Power Supply Replacement Procedure (cont’d)
6.) Loosen and remove the 5 screws that secure the DC-to-DC Power Supply to the cage assembly
(see Figure 8-178 below).
7) Disconnect both of the Power Supply cables from the DC-to-DC Power Supply, as shown below:
HVPS-to-RFI (shown left in Figure 8-179) and HVPS-to-BEP (shown right in Figure 8-179).
8) Using both hands, carefully lift the DC-to-DC power supply upwards and remove it from the cage
assembly, as shown in Figure 8-180 below.
Figure 8-178 Loosening the DC-to-DC PS Securing Screws
Figure 8-179 Disconnecting the DC Power Supply Cables
Figure 8-180 Removing the DC-to-DC Power Supply