GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-225
PRELIMINARY
19. Navigate to the TCPIP tab and press the Set button on
the control panel.
Connectivity dialog box - TCPIP tab activated:
20. Navigate to the Save Settings button and press the Set
button on the control panel.
A message is displayed prompting for system re-boot:
21. Click OK and re-boot the system.
22. If this is a new software version, an adhesive Rating Plate
is supplied with the software CD.
Adhere the plate in the appropriate position at the rear of
the system.
23. Proceed to perform the following functionality tests:
• “Software Configuration Checks” on page 4-45
• “Calibration Tests” on page 7-27
• “Full System Test” on page 7-11
Table 8-3 Vivid™ S5 N and Vivid™ S6 N Software Re-Installation (cont’d)
Step Instruction Expected Result