1494 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
39 Obsolete and Discontinued Commands
:MEASure:VTIMe (see
page 1531)
:MEASure:YATX (see
page 640)
With :MEASure:YATX, there is a
command for installing the
measurement on the
oscilloscope's display.
:MEASure:VTIMe had no
command, only a query.
:MTESt:AMASk:{SAVE | STORe}
(see page 1532)
:SAVE:MASK[:STARt] (see
page 834)
:MTESt:AVERage (see
page 1533)
:ACQuire:TYPE AVERage (see
page 318)
:MTESt:AVERage:COUNt (see
page 1534)
:ACQuire:COUNt (see
page 304)
:MTESt:LOAD (see page 1535) :RECall:MASK[:STARt] (see
page 816)
:MTESt:RUMode (see
page 1536)
:MTESt:RMODe (see
page 684)
:MTESt:RUMode:SOFailure (see
page 1537)
:MTESt:RMODe:FACTion:STOP
(see page 688)
:MTESt:{STARt | STOP} (see
page 1538)
:RUN (see page 291) or :STOP
(see page 295)
:MTESt:TRIGger:SOURce (see
page 1539)
:TRIGger Commands (see
page 1295)
There are various commands
for setting the source with
different types of triggers.
:PRINt? (see page 1540):DISPlay:DATA? (see
page 412)
:SAVE:IMAGe:AREA (see
page 1542)
none
:TIMebase:DELay (see
page 1546)
:TIMebase:POSition (see
page 1284) or
:TIMebase:WINDow:POSition
(see page 1291)
TIMebase:POSition is position
value of main time base;
TIMebase:WINDow:POSition is
position value of zoomed
(delayed) time base window.
:SBUS<n>:LIN:SIGNal:DEFinitio
n (see page 1543)
none
:SBUS<n>:SPI:SOURce:DATA
(see page 1544)
:SBUS<n>:SPI:SOURce:MOSI
(see page 1083)
:SYSTem:MENU (see
page 1545)
:DISPlay:MENU (see
page 420)
No change in behavior.
Obsolete Command Current Command Equivalent Behavior Differences