ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 3 - System Setup  3-99
PRELIMINARY
9.) Select the appropriate settings to configure Dicom Storage (e.g. Name: DICOM Storage 1, etc. as 
shown in Figure 3-113).
10.)Under DICOM SR Settings, activate the Allow SR checkbox, if applicable.
11.)When done, click OK to close the dialog box.
EchoPAC / Image Vault sites:
1.) Access the Configuration > Connectivity settings and click on the Dataflow tab.
2.) Select the dataflow named Remote Archive - Remote HD - see Figure 3-114.
3.) Under Selected devices > Inputs, select Remote Database.
4.) Click on Properties.
Figure 3-114   Selecting Required Dataflow