ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
7-136 Section 7-7 - Automatic Error Log
PRELIMINARY
7-7-2 Extracting Logs (cont’d)
1.) Perform step 1 and step 2 on page 7-132.
2.) Click the Extensive Log button (refer to step 7-94 on page 7-133).
The Extensive Logging dialog opens - see step 7-96, below.
3.) Flag the appropriate checkboxes to activate the categories for which detailed analysis is required
(refer to the recommendations in Table 7-5).
NOTE: The impact of activating a checkbox would be to generate larger, detailed log files.
After usage, make sure to return to the default settings (unflagged) to conserve the size of the
log files. After 7 days, the system will automatically return to the default settings.
4.) When done, click OK.
The Extensive Logging dialog closes.
Figure 7-96 Extensive Logging Dialog