4.4 Judging Measurement Results
94
The comparator decision mode can be set as one of the following:
Deviation Percent (%) Setting (p. 91)
HI
IN
IN
LO
upper limit value[%]
reference value[]
lower limit value[
%]
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit values.
The measurement values are displayed in deviations (%) from the ref-
erence value.
Percent (%) Setting (p. 89)
HI
IN
IN
LO
upper limit value[%]
reference value[]
lower limit value[%]
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit values.
The measurement values displayed are the same as those of the mea-
surement parameters.
Absolute value (ABS) setting (p. 88)
HI
IN
LO
upper limit value
lower limit value
Set absolute values for the upper limit and lower limit values of
the measurement parameters.
The measurement values displayed are the same as those of the mea-
surement parameters.
• If the power is turned off in BIN measurement mode, the mode will be BIN measurement
mode when the instrument starts the next time you turn the power on.
• For a BIN number that does not require a BIN judgment, set the upper and lower limit val-
ues to OFF.
• The measurement conditions that are used when normal measurement is performed are
inherited as is for the measurement conditions when BIN is performed.
*2: The following equation is used to calculate the % value.
% =
|reference value|
measurement value - reference value
×100
*1: The following equation is used to calculate the comparison upper limit value and comparison lower limit value.
(In the case of the comparison lower limit value, if a value that is lower than the reference value is set, the minus
(-) sign is required for the percentage setting value.)
Upper limit comparison value (Lower limit comparison value) = reference value+ |reference value| ×
100
percentage set value