7.3 Compensating Values to Match Reference Values (Load Compensation)
293
7
Chapter 7 Error Compensation
Set the compensation frequency.
9
When the load compensation is valid for the set
measurement conditions, ON appears on the
LOAD parameter in the Initial Screen.
When the same compensation frequency has been set
to multiple load compensation groups, only the group
with the smallest number will be valid.
When the current measurement frequency does not
agree with the load compensation frequency, the load
compensation will be invalid and ON will not appear.
LCR Initial Screen
2. Use the numeric keypad to enter the compensation fre-
quency.
Settable range: DC, 4.00 to 5.0000 MHz
3. Press a unit key to confirm the setting.
4. Press to close the setting screen.
Compensation Frequency Setting
• When performing load compensation for DC resis-
tance measurement: Press .
• If you make a mistake during input:
press to cancel the input and start again.
• When you want to cancel input:
Press to close the compensation fre-
quency setting screen.
Compensation Frequency Setting
1. Press .