4.2 Setting Basic Settings of Measurement Conditions
60
Depending on the measurement signal level, in some cases it is possible to damage the sample which is
being tested by applying to it a voltage or a current greater than its rated value.
For this reason, set a limit value to restrict the voltage that can be applied to the test sample or current that
can flow to the test sample.
4.2.7 Setting the Voltage/Current Limit
When open circuit voltage (V) or
constant voltage (CV) is set:
Set the current limit.
When constant current (CC) is set:
Set the voltage limit.
Procedure
LCR Initial Screen
• The present situation can be checked using the monitor display.
• The monitor display is different for V, CV, and CC.
First set the measurement signal level, and thereafter set the voltage or current limit.
The setting for voltage or current limit changes automatically to current or voltage limit,
according to the present measurement signal level setting.
See "4.2.2 Setting the Measurement signal level" (p. 43)
1
LCR Basic Settings
3
When the measurement signal level is a
voltage (V, CV)
Current Limit Setting
When the measurement signal level is a
current (CC)
Voltage Limit Setting