7.3 Compensating Values to Match Reference Values (Load Compensation)
289
7
Chapter 7 Error Compensation
• When all of the compensation conditions match for load compensation, compensation is
performed under the optimal conditions.
• If the current measurement frequency and compensation frequency do not match, an
error like the following is displayed in the initial screen.
• If a condition other than the compensation frequency does not match, compensation is
performed but a warning like the following is displayed in the initial screen.
• When the OPEN or SHORT compensation is valid, the load compensation is performed
for Z and processed by the OPEN or SHORT compensation.
• In acquiring load compensation data (i.e., reference sample measurement), the OPEN/
SHORT compensation settings, that were defined before entry into the Load Compensa-
tion Screen, are valid.
• If the setting of the low Z high accuracy mode is changed, the compensation value
becomes invalid.