7.3 Compensating Values to Match Reference Values (Load Compensation)
289
7
Chapter 7 Error Compensation
When all of the compensation conditions match for load compensation, compensation is
performed under the optimal conditions.
If the current measurement frequency and compensation frequency do not match, an
error like the following is displayed in the initial screen.
If a condition other than the compensation frequency does not match, compensation is
performed but a warning like the following is displayed in the initial screen.
When the OPEN or SHORT compensation is valid, the load compensation is performed
for Z and processed by the OPEN or SHORT compensation.
In acquiring load compensation data (i.e., reference sample measurement), the OPEN/
SHORT compensation settings, that were defined before entry into the Load Compensa-
tion Screen, are valid.
If the setting of the low Z high accuracy mode is changed, the compensation value
becomes invalid.