2.2 Pre-Operation Inspection
28
2.2 Pre-Operation Inspection
Do not use the instrument if damage
is found, as electric shock or short-
circuit accidents could result. Con-
tact your dealer or Hioki representa-
tive.
Metal Exposed
Is the power cord insulation torn, or is
any metal exposed?
1
No Metal Exposed
Before using the instrument for the first time, verify that it operates normally to ensure that no damage
occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative.
Peripheral Device Inspection
Is the insulation on a measurement
cable torn, or is any metal exposed?
Metal Exposed
If there is any damage, measured
values may be unstable and mea-
surement errors may occur. Replace
the cable with an undamaged one.
No Metal Exposed
If damage is evident, request repairs.
Yes
Is damage to the instrument evident?
Instrument Inspection
When turning power on
Does the self-test screen appear (model
no., version no.)?
No
2
The power cord may be damaged, or
the instrument may be damaged in-
ternally. Request repairs.
Does the Measurement screen appear
after self-test?
No
Yes
The instrument may be damaged in-
ternally. Request repairs.
See "14.1 Inspection, Repair and
Cleaning" (p. 403)
An error indication
occurs (ERR)
Yes
Inspection complete