EasyManua.ls Logo

Hioki IM3570 - 14.3 Error display

Hioki IM3570
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13.3 About Measurement Times and Measurement Speed
399
13
Chapter 13 Specifications
Measurement times differ depending on the measurement conditions. Refer to the following values.
When there is an external trigger, the corresponding time shown in the table below is added to the analog
measurement signal.
* Low Z high accuracy mode is only valid when the frequency is between 4 Hz and 100 kHz.
13.3 About Measurement Times and
Measurement Speed
All of the values are reference values. Note that they may differ depending on the conditions of use.
Analog measurement signal (INDEX)
FAST MED SLOW SLOW2
DC 1 ms+0.6 ms 10 ms+0.6 ms 100 ms+0.6 ms 400 ms+0.6 ms
4.00 Hz to 19.99 Hz Tf+0.6 ms 2 × Tf+0.6 ms 3 × Tf+0.6 ms 4 × Tf+0.6 ms
20.00 Hz to 100.00 Hz Tf+0.6 ms 2
× Tf+0.6 ms 8 × Tf+0.6 ms 64 × Tf+0.6 ms
100.01 Hz to 999.99 Hz Tf+0.6 ms 4 × Tf+0.6 ms 16 × Tf+0.6 ms 128 × Tf+0.6 ms
1.0000 kHz to 3.0000 kHz Tf+0.2 ms 8
× Tf+0.2 ms 32 × Tf+0.2 ms 256 × Tf+0.2 ms
3.0001 kHz to 10.000 kHz Tf+0.2 ms 16 × Tf+0.2 ms 64 × Tf+0.2 ms 512 × Tf+0.2 ms
10.001 kHz to 30.000 kHz 4
× Tf+0.2 ms 64 × Tf+0.2 ms 256 × Tf+0.2 ms 2048 × Tf+0.2 ms
30.001 kHz to 100.00 kHz 8 × Tf+0.1 ms 128 × Tf+0.1 ms 512 × Tf+0.1 ms 4096 × Tf+0.1 ms
100.01 kHz to 300.00 kHz 32
× Tf+0.1 ms 512 × Tf+0.1 ms 2048 × Tf+0.1 ms 16384 × Tf+0.1 ms
300.01 kHz to 1.0000 MHz 64 × Tf+0.1 ms 1024 × Tf+0.1 ms 4096 × Tf+0.1 ms 32768 × Tf+0.1 ms
1.0001 MHz to 5.0000 MHz 400
× Tf+0.1 ms 6400 × Tf+0.1 ms 25600 × Tf+0.1 ms 204800 × Tf+0.1 ms
Tolerance: ±5% ±0.2 ms Tf [s]=1 ÷ measurement frequency
The DC time is the time when DC adjustment is OFF. The time is about twice as long when DC adjustment is ON.
Range Level Frequency Low Z High Accuracy Mode DC Bias Time
100 m
, 1 All levels DC, 4 Hz to 999.99 Hz ON OFF 5 ms
100 m, 1 All levels 1 kHz to 5 MHz * ON OFF 1 ms
100 m
, 1 All levels ALL frequency * ON ON 1 ms
100 m, 1 All levels DC, ALL frequency OFF OFF 1 ms
10
0.005 V to 0.1 V DC, ALL frequency OFF OFF 1 ms
10 0.101 V to 1 V DC, 4 Hz to 1 MHz OFF OFF 1 ms

Table of Contents