Hioki IM3570 - 8.3 Self Checks (Self Diagnosis)

Hioki IM3570
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.3 Compensating Values to Match Reference Values (Load Compensation)
299
7
Chapter 7 Error Compensation
If you press , all settings are cancelled and you can start again from the compensation fre-
quency settings.
Press to acquire the current measurement conditions (frequency, range, measurement sig-
nal type and level, and DC bias settings) and use them as the load compensation conditions.
When You Want to Reset All Settings
When You Want to Use the Current Measurement Conditions as the
Load Compensation Conditions
Load Compensation Setting
Load Compensation Setting
When the measurement conditions are acquired with , ("Setting of
Parameter to Use for Reference Value" (p. 297)) is initialized to Z-
.