5.9 Application Settings
228
You can select whether to enable or disable trigger input from the EXT I/O during measurement (during EOM
(HI) output after trigger is received). Furthermore, you can also select either the rising edge or falling edge as
the valid edge of trigger input from the EXT I/O.
See "11.2 Timing Chart" (p. 369)
5.9.5 Enabling Trigger Input for during Measurement
and Setting the Valid Edge of Trigger Input
Procedure
Application Settings
1
Analyzer Initial Screen
3
I/O Trigger Setting
4
Press to close the setting screen.
Press .
2
Application Settings
Sets the falling edge as the valid edge of trigger
input.
Sets the rising edge as the valid edge of trigger
input.
Disables trigger input from the EXT I/O during
measurement (during EOM
(HI) output after
trigger is received).
Enables trigger input from the EXT I/O during
measurement (during EOM
(HI) output after
trigger is received).
Select the I/O trigger function setting.