6.5 Setting Application Settings
271
6
Chapter 6 Continuous Measurement Function
Measurement can be halted when the judgment result is FAIL (HI/ LO/ OUT OF BINS) and the judgment func-
tion for the measured panel (comparator, BIN function) is enabled. Continuous measurement is halted regard-
less of the settings when the measurement result yields a contact check error or HIGH-Z reject error in low Z
high accuracy mode.
6.5.3 Halting Continuous Measurement at a FAIL judg-
ment result
Procedure
Continuous Measurement Initial Screen Application Settings
1
Press .
2
Application Settings
Turn the setting to halt continuous measure-
ment on or off.
Performs continuous measurement for all
panels, regardless of the judgment result.
Halts continuous measurement when the
judgment result is FAIL.
3
Continuous Measurement Halt Method Setting
4
Press to close the setting screen.