13.2 Measurement Range and Accuracy
398
Guaranteed Accuracy Range
The guaranteed accuracy range differs depending on the measurement frequency, measurement signal level,
and measurement range.
Range DC
4.00 Hz to
99.99 Hz
100.00 Hz to
999.99 Hz
1.0000 kHz to
10.000 kHz
10.001 kHz to
100.00 kHz
100.01 kHz to
1.0000 MHz
1.0001 MHz to
5.0000 MHz
100 M 0.1 V to 2.5 V 0.101 V to 5 V 0.501 V to 5 V
10 M
0.1 V to 2.5 V
0.050 V to 5 V 0.101 V to 5 V 0.501 V to 5 V
1 M 0.050 V to 5 V 0.101 V to 5 V 0.501 V to 1 V
100 k 0.050 V to 5 V 0.101 V to 1 V
30 k
0.05 V to 1 V
10 k
0.005 V to 5 V3 k
1 k
300
10
1 0.005 V to 5 V *2 0.101 V to 5 V 0.501 V to 1 V
100 m
0.1 V to 2.5 V *1
0.101 V to 5 V *3 0.501 V to 5 V*3
The above voltages are the voltage settings equivalent to when V mode.
*1 The guaranteed accuracy is 10 m or higher
*2 The guaranteed accuracy for when DC bias is 0.101 to 5 V.
*3 The guaranteed accuracy for when DC bias is 10 m or higher and 1.001 to 5 V.
The above measurement specification was determined using a 1.5C-2 V coaxial cable with
an established cable length for the unit.
Using a cable other than a 1.5C-2 V, or a cable that not an established length for the unit in
question increases the chance of measurement inaccuracy. A large capacitance between
the H terminal and grounding capacitance (GND) or the L terminal and GND may result in
measurement inaccuracy. Please set the GND to 10 pF or less.