4.5 Setting Application Settings
112
• If a detection level error is detected, the following comment is displayed at the top of the
screen.
• The detection level error is reflected in the response result of
:MEASure?, :MEMory?.
See Description of communications commands on the included CD (:MEASure?, :M EMory?)
Furthermore, it is output as a measurement error to the ERR terminal (10-pin) of the EXT I/O.
See "Connector Type and Signal Pinouts" (p. 364)
• Even in a normal contact state, an error may be output due to the influence of external
noise. (It is possible to judge external noise.)
Use the shielding process as a measure against external noise.
• Chattering monitoring is not performed during offset measurement when DC adjustment is
enabled.
See "4.3.3 Setting the DC Adjustment Function" (p. 75)