11.2 Timing Chart
371
11
Chapter 11 External Control
The higher the measurement frequency, the shorter the time that INDEX and EOM are high (off). When the
high (off) time is too short due to characteristics of the input circuit, the instrument can be configured to
maintain the low (on) state for a preset time once EOM
changes to low (on) before reverting the signal to
high (off) after the completion of measurement.
When trigger input is received at EOM
:LOW and INDEX:LOW, the signal transitions to high (off) when mea-
surement starts.
Setting the INDEX
and EOM output method
See "4.5.8 Setting the EOM Output Method" (p. 119)
Description of communications commands on the included CD (:IO:EOM:MODE)
Setting the pulse width for which low (on) EOM is held
See "4.5.8 Setting the EOM Output Method" (p. 119)
Description of communications commands on the included CD (:IO:EOM:PULSe)
(Start-of-measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
When HOLD
When PULSE
When HOLD
When PULSE
T: Pulse width