• ERR transitions to LOW each time measurement completes if a measurement error occurs.
• In the continuous measurement screen, comparator result output signals other than AND and
panel load signals (LD-VALID
, LD0 to LD6) cannot be used.
See: "Chapter 6 Continuous Measurement Function" (p. 265)
• Whether the judgment results of comparator measurement are reset at the time of the mea-
surement start signal or updated at the point in time when measurement ends can be selected
on the instrument or by a communication command.
See: "Setting the Delay Time from the Output of Comparator and BIN Judgment Results until Output of EOM
(LOW) and Resetting Judgment Results" (p. 115),(p. 226)
Description of communications commands on the included CD (
:IO:RESult:RESet)
• For other timing chart times, refer to "11.2.1 LCR Measurement" (p. 369).