4.3 Setting DC Resistance Measurement
76
4
Press to close the setting screen.
3
DC Adjustment Setting
Select ON/ OFF for DC adjustment.
Obtains the offset value with the CALIB
signal of EXT I/O or a communication
command (:
DCResistance:ADJust
:DEMand
).
Obtains the offset value for each measure-
ment.
Since the transient response time becomes long as a result of the impedance value of the measure-
ment object, set a long DC delay time.(p. 82)
When DC adjustment is ON, the measurement time becomes about twice as long because time is
required both for normal measurement and for offset measurement.
DC adjustment measurement is performed in the following cases even if DC adjustment is set to OFF.
When measurement is first performed after the power is turned on
When the DC delay time is changed
When the measurement range is changed
When low Z high accuracy mode is turned ON/OFF
When the measurement signal level is changed