4
Press to close the setting screen.
3
DC Adjustment Setting
Select ON/ OFF for DC adjustment.
Obtains the offset value with the CALIB
signal of EXT I/O or a communication
command (:
DCResistance:ADJust
:DEMand
).
Obtains the offset value for each measure-
ment.
• Since the transient response time becomes long as a result of the impedance value of the measure-
ment object, set a long DC delay time.(p. 82)
When DC adjustment is ON, the measurement time becomes about twice as long because time is
required both for normal measurement and for offset measurement.
• DC adjustment measurement is performed in the following cases even if DC adjustment is set to OFF.
• When measurement is first performed after the power is turned on
• When the DC delay time is changed
• When the measurement range is changed
• When low Z high accuracy mode is turned ON/OFF
• When the measurement signal level is changed