Operation Manual SUPREMA
11. Technical Data
11.2 Module Data
11.2.1 MCP 10 Central Processing Unit
Order No.: 10022752
Function
- Monitoring and control of all system functions, signal evaluation for up to 256 sampling
points
- actuation of up to 512 relay driver outputs
- parameter storage
Technical Data
Operating voltage feed (3 x 24 VDC): 14 ... 32 VDC
Operating current: 75 mA
Temperature range: 5 °C ... 40 °C
Humidity: 0 ... 90 % relative humidity
noncondensing
Plug connector: 96-channel VG terminal strip
(pre-mating contacts for power supply)
Dimensions: 100 x 160 mm
Weight: 120 g
11.2.2 MDA 10 Data Acquisition Unit
Order No.: 10022688
Function
- reads in the measurement values from the MAI10 module, processes the
measurement signals, calculates mean values
- monitors 2 x power supply voltages, 1 x battery voltage
Technical Data
Operating voltage feed (3 x 24 VDC): 14 ... 32 VDC
Operating current: 95 mA
Temperature range: 5 °C ... 40 °C
Humidity: 0 ... 90 % relative humidity
noncondensing
Plug connector: 96-channel VG terminal strip
(pre-mating contacts for power supply)
Dimensions: 100 x 160 mm
Weight: 108 g
11.2.3 MDO 10 Display + Operation Unit
Order No.: 10022686
Function
- 240 x 64 pixel backlit graphics display
- system operation via cursor field