REV 0, June 1005 Page 245
Operation Manual SUPREMA
11. Technical Data
11.2.19 MAT 10 Analog Terminal Unit
Order No.: 10015759
Function
- connecting terminals for remote measuring heads
Technical Data
Number of measuring head connections: 8
Number of terminals per sampling point: 5
Allowable wire cross section 1,5 mm
2
Temperature range: 5 °C ... 40 °C
Humidity: 0 ... 90 % relative humidity
noncondensing
Dimensions: 125 x 50 mm
Weight: 155 g
11.2.20 MSP 10 System Power Unit
Order No.: 10020340
Function
- rack power supply, 150 VA
Technical Data
Operating voltage, feed: 85 ... 265 VAC
Maximum operating current: 2 A
(100 VAC Input)
; 1 A
(200 VAC Input)
Maximum switch on current: 50 A at 230 V (cold start)
Power correction factor: according to EN 61000-3-2
Interference emissions: according to EN 55011/EN 55022-B
Output voltage: 24 VDC
Maximum output current: 6.5 A
Temperature range: 5 °C ... 40 °C
Humidity: 0 ... 90 % relative humidity
noncondensing
Dimensions: 65 x 92 x 198 mm
Weight: 850 g