REV 0, June 1005 Page 239
Operation Manual SUPREMA
11. Technical Data
11.2.7 MPI 10 WT100 Passiv Input Unit
Order No.: 10021028
Function
- power source for passive sensors (24 VDC)
- supply module for WT sensors
- sensor current, 100-400 mA
- prepares the Ux bridge signal
Technical Data
Operating voltage feed: 19.2 ... 32 V
Operating current: 460 mA, maximum
Maximum load: 36 ohms
(at operating voltage of 19.2 V and
sensor current of 400 mA)
Setting range, constant current: 180 ... 400 mA
Setting range, zero point: ± 130 mV (Ux)
Setting range, sensitivity: 4 ... 190 mV (Ux)
Temperature range: 5 °C ... 40 °C
Humidity: 0 ... 90% relative humidity
noncondensing
Dimensions: 46 x 24 mm
Weight: 12 g
11.2.8 MPI 10 WT10 Passiv Input Unit
Order No.: 10024279
Function
- power source for passive sensors (24 VDC)
- supply module for WT sensors
- sensor current, 100 ... 400 mA (depending on components installed)
- prepares the Ux bridge signal
Technical Data
Operating voltage feed: 19.2 ...32 V
Operating current: 460 mA, maximum
Maximum load: 28 ohms
(at operating voltage of 19.2 V and
sensor current of 400 mA)
Setting range, constant current: 180 ... 400 mA
Setting range, zero point: ± 50 mV (Ux)
Setting range, sensitivity: 5 ...27 mV (Ux)
Temperature range: 5 °C ... 40 °C
Humidity: 0 ... 90 % relative humidity
noncondensing
Dimensions: 46 x 24 mm
Weight: 12 g

Table of Contents

Related product manuals