REV 0, June 2005Page 244
Operation Manual SUPREMA
11. Technical Data
Humidity: 0 ... 90 % relative humidity
noncondensing
Maximum load: 500 ohms
Output current 4 ... 20 mA: 0 ... 24 mA
Dimensions: 100 x 160 mm
Weight: 127 g
11.2.18 MIB 20 Interconnection Board
Order No.: 10032525
Function
- rack rear-panel wiring for 3 x MCP, 2 x MDA, 8 x MAI, 2 x MGO modules
- power supply for all modules (INT, EXT, BAT)
- connection for 3 x 24 VDC feed, screw terminals
- provides uninterruptible 24-V-system power
- data transfer between the modules over a CAN or SPI bus
- 2 system error relay, 1 change-over contact, 3 connecting terminals
- DII switch for CAN rack ID, CAN bit rate, CAN bus terminating resistor
- 7 “dedicated” slots for 3 x MCP, 2 x MDA, and 2 x MDO modules
- 7 “free” slots for MAI, MPI, MLI, MBI, MGO, and MGI modules
- electric connection for the inserted modules
- connection modules (MST, MAT, etc.) are plugged into the rear of the MIB
Technical Data
Operating voltage feed: 19.2 ... 32 V
Maximum allowable operating current:
feed (+24 v) 20 A
(GND): 32 A
Feed connector cross section: 4 mm
2
, flexible
6 mm
2
, rigid
Power supply terminals: EXT, INT, BAT
Setting elements: 10-channel DIL switch
(rack no., CAN bit rate)
Contact load capacity,
system error relay 3 A
Temperature range: 5 °C ... 40 °C
Weight: 0 ... 90 % relative humidity
noncondensing
Abmessungen: 377 x 128 mm
Weight: 650 g