CAPINTEC, INC. CAPTUS
®
4000e
August 15 BIOASSAY 11-7
To abort the procedure, select the Yes button or press the Y key. Figure 11-5 will
re-appear. Select the Start Procedure button to restart the measurement or the
Exit button to return to the Bioassay Setup screen (Figure 11-2).
vii. If more than one isotope was selected for the Efficiency measurement, the
program will return to Figure 11-8 and prompt to measure the next isotope.
viii. After all measurements are completed, the screen shown in Figure 11-10 will
appear. The results in the lower part of the screen show all measured counts and
calculated values for activity and cross contamination.
Figure 11-10
ix. To print the results on the screen, select the Print button or press Alt+P.
x. To exit and return to the main Bioassay Setup screen (Figure 11-2), select the
Exit button or press Alt+E.
b. To directly enter the Efficiency of a selected isotope or Cross Contamination ratio,
follow the steps below:
i. Select the Enter Efficiency button or press Alt+E. The screen appears as shown
in Figure 11-11.