5.4 Segment Sweep
170
As shown in the figure above, when a segment sweep is performed, the following items can be set indi-
vidually for each segment.
Sweep range
Number of sweep point
Setting method of sweep points
Measurement signal level
Range
Average
Measurement speed
Point delay
When segment sweep
Sweep Settings Segment 1 Segment 2 Segment 3
Sweep parameter Frequency Frequency Frequency
Sweep Range 1.0000kHz to 5.0000kHz 10.000kHz to 100.00kHz 100.00 kHz to 5.0000MHz
Number of sweep point 201 points 201 points 399 points
Setting method of sweep
points
Log Log Linear
Measurement signal type Open circuit voltage level Open circuit voltage level Open circuit voltage level
Measurement signal level 1.000V 1.500V 0.500V
Range AUTO 3 k
AUTO
Average 10 times 3 times OFF
Measurement speed FAST MEDIUM SLOW
Point delay 0.0000 s 0.0005 s 0.0000 s
Segment1 Segment2 Segment3
The following settings are common to all segments.
Measurement parameter
•Sweep parameter
Measurement signal type
Trigger
Trigger delay
DC bias level