5.7 Judging Measurement Results (Comparator Function)
199
5
Chapter 5 ANALYZER Function
Press .
Comparator Setting
4
Press to confirm the setting.
Judgment Parameter Setting
5
Select the parameter to judge.
Sets the upper and lower limit values and
judges the measurement results for the
first measurement parameter.
Sets the upper and lower limit values and
judges the measurement results for the
second measurement parameter.
Sets the upper and lower limit values and
judges the measurement results for both
the first and second parameters.
If a parameter is changed, the comparator setting prior to the change becomes invalid, and the
comparator function is set to .
If a measurement point is changed, the comparator setting prior to the change becomes invalid,
and the comparator function is set to .
See: "5.3.1 Setting Sweep Points" (p. 143)