7.2 Short Circuit Compensation
286
The compensation values from last time are dis-
played in a confirmation screen.
(If compensation has never been performed, the
compensation values become 0.)
5
Spot Compensation Value Setting
6
Compensation starts.
Compensation value acquisition time:
Approximately 5 seconds
Performing Short Compensation
Check that the measurement cable is in a short-circuit
state.
Press
.
When you want to cancel compensation:
Press .
Compensation is cancelled and the compensation
screen is redisplayed.
(The short circuit compensation values from last time
remain.)
When you do not want to acquire the compen-
sation values: Press .
The setting screen is redisplayed, and the compensa-
tion values from last time become valid.
Spot Compensation
7
When compensation failed: (p. 287)
If compensation ends normally, the effective resis-
tance and reactance are displayed.
The possible compensation range is 1 k
or less for
impedance.
Use or to check the effective resistance
and reactance of each compensation point.
8
Press to close the setting screen.
Indicates the
compensation
numbers.
Indicates the
measurement
frequencies.
Indicates the compensation results.
(Effective resistance, reactance)