7.2 Short Circuit Compensation
287
7
Chapter 7 Error Compensation
When an error message appears and compensation has stopped, short circuit compensation is
turned off.
Check the following points before starting the short circuit compensation process again:
• Check that the test cables are properly connected.
• Check that the test cables are properly shorted together with the shorting bar. (Short circuit compensa-
tion cannot be performed while any test sample is connected to the test cables.)
• Check that the test leads are arranged as closely as possible to their configuration in which measure-
ment will be performed.
• During the compensation process, be sure not to disturb the test cables or to move your hand near them.
Press in Step
3 of [Short Compensation Setting] (p. 282) to make the compensation data for up
until now invalid and return to the [Initial screen].
When Short Compensation Failed
When You Want to Make Short Circuit Compensation Data Invalid
Short Compensation Setting
The compensation values that are stored internally are not cleared by the operation
described above. When ALL or SPOT is selected, the stored compensation values can be
used.